In the present work, a mathematical model for the impedance response of glucose-oxidase based electrochemical biosensors has been developed [3]. The complicated coupling between thirteen homogeneous reactions and three heterogeneous reactions has been considered in the model. The homogeneous reactions included anomerization between α-D-glucose and β-D-glucose and four reversible enzymatic catalytic reactions transforming β-D-glucose and oxygen into gluconic acid and hydrogen peroxide, pH-dependent enzymatic activity and a biological buffer system. The electroactive hydrogen peroxide was considered to be reversibly oxidized or reduced at the electrode.
The mathematical model is solved numerically by using the finite-difference method and Newmanâs BAND algorithm [4]. The model demonstrates how the coupled non-linear homogeneous reactions effect the diffusion impedance, which has broadened the scope of the Gerischer impedance. The model can be used to explore the influence of various system parameters on limiting current, reaction profiles, and diffusion impedance. The system parameters, including interstitial glucose concentration, oxygen concentration, active enzyme concentration, diffusion coefficients, reaction rate constants and layer thickness, are related to various sensor working conditions such as body sugar level, inflammation, sensor degradation and sensor design.
References
Acknowledgement
The support of Medtronic Diabetes (Northridge, CA) and Andrea Varsavsky, program monitor, is gratefully acknowledged.