2016 AIChE Annual Meeting
(237g) Gas Phase Composition Measurements Using Ultrasonics
Considering all of the factors which could cause inaccuracy, the standard uncertainty of this SOS measurement is estimated in this instrument to be < ±0.01 m/s. A temperature-speed of sound-composition database has been constructed for 20 industrially important air-vapour and several binary gas systems over a temperature range from 5 to 55 °C. The SOS across the temperature range of 5 to 55 °C and the gas composition has also been independently determined using a highly accurate gas buoyancy method. The average deviation in accuracy between the predicted and the measured composition for 15-55 °C is typically < 0.5%, and for specific test temperatures can be <0.1%. Typical precision is 1: 300000. In comparison with traditional measurement techniques, this technique is not only robust but also has the potential to be faster, simpler, more cost-effective, more precise and more accurate.