(686a) Simultaneous Surface Manipulation and Sensing in a Biosensor Using a Hexagonal Saw Device
Authors
In this device, the Rayleigh acoustic waves stress the bonds between the sensing film and analyte forcing only the analyte with higher affinity for the sensing film to stay bound, while the SH SAWs are used for sensing. Additionally, the acoustic waves work to efficiently mix the liquid samples flowing through the micro-channels of the micro-sensor system, reducing the effects of diffusion-limited processes.
Results from using a sensing film of anti-mouse IgG covalently bound to the sensor-surface and mouse IgG as the analyte in buffer solution have shown improved sensor response, determined using fluorescent microscopy. Manipulation of liquid samples was achieved by strongly exciting the piezoelectric substrate with power levels of ~12 mW which is significantly greater than the 1 mW used for sensing. The larger electrical power creates an acoustic wave via piezoelectric coupling that can physically force loosely bound species from binding sites, reducing noise that can lead to inaccurate measurements.