(677g) A New Technology for Pulmonary Drug Delivery
Authors
Our approach is to utilize our new ultrafine powder handling technology to meter and deliver the drugs, so that these ultrafine drugs can be used alone, without the "aids" of coarse powder or liquid (excipients). The first key achievement of the new technology is the ability to accurately dispense a tiny quantity (in the order of 0.02 to 0.5 milligrams per dose) of the ultrafine drugs (< 5 microns). The second key achievement is the development of a more effective inhaler that gives much higher delivery efficiency (> 50%) than those currently on the market. This research project, with several patents granted, has produced a combination of new drug handling processes and new inhalers that are likely to eventually set an industry standard for pulmonary drug delivery in the near future.
A rotating fluidized bed powder dispensing device was developed to dispense the extrafine drug particles into a novel dry powder inhaler (DPI). Tests were carried out with various 1-5 micron powders such as salbutamol sulphate, insulin, lactose, and mannitol. The dispensing device is designed to permit filling of up to 60 doses per minute, with weights ranging from 20 micrograms - 2000 micrograms metered into a multi-dose blister pack, integrated into the DPI.
The patented rotating fluidized bed powder dispensing device has a rotating porous cylindrical chamber with alternating distributor and ceiling, and a flexible nozzle for additional agitation. It is designed to prevent the agglomeration of extrafine particles in the absence of any large excipient particles. With such an arrangement, the extrafine particles can be fluidized inside the chamber to produce a uniform gas-particle suspension which can be withdrawn from the fluidized chamber as a gas-particle plume. At the withdrawal port, up to 60 doses per minute (20 to 2000 micrograms) can be metered into a multi-dose blister pack.
The breath-activated, excipient-free dry powder inhaler was designed to facilitate delivery of drug powders deep into human lungs. The Andersen cascade impactor, Next Generation Impactor and TSI particle size distribution analyzer were used to evaluate the performance of the DPI in conjunction with salbutamol sulphate, insulin, lactose and mannitol.
Testing revealed that the rotating fluidized bed powder dispensing device can provide a dilute gas-solid suspension with more than 90% of the particles present in discrete (non-aggregated) suspended form. For example, results show that the majority of insulin particles in the plume exiting the fluidized dispensing device are under 3 micrometer with a narrow particle size distribution and only one peak. TSI sizing clealy indicates that the starting powder can be broken into single particles efficitively. The dose uniformity of each powder filled by the dispensing device was tested using disks with different dosage sizes (20 - 2000 micrograms). The deviations among the doses in the same disk and between the disks under the same condition were very small - the relative standard deviation (within the disk) of the filled weights into the different size dosage disk was less than 5%. The new inhaler was also found to be very effective, much more than those currently on the market. In-vitro tests have shown that the newly designed DPI exhibits a high fine particle fraction ( < 4.7 microns) of approximately 60%-80%, using both Andersen cascade impactor and Next Generation Impactor.
Testing suggests that the combination of the new breath-activated and excipient-free dry powder inhaler and the rotating fluidized bed powder dispensing device can efficiently dispense and deliver extremely small quantities (as low as 20 micrograms) of extrafine drug particles with the RSD below 8%.